The absolute measurement of low power at 3000 Mc/s
作者:
R.Street,
期刊:
Proceedings of the IEE - Part III: Radio and Communication Engineering
(IET Available online 1949)
卷期:
Volume 96,
issue 41
页码: 237-242
年代: 1949
DOI:10.1049/pi-3.1949.0051
出版商: IEE
数据来源: IET
摘要:
A general method for the calibration of milliwattmeters operating in the centimetric-wave region is described. Results of measurements at 3000 Mc/s on a thermistor bridge milliwattmeter are given and it is shown that within the present limits of experimental error (estimated as±4%) this instrument gives true power indications. The powers indicated by other forms of milliwattmeter, for equal power inputs, have been determined and the results are summarized. It is concluded that a number of different instruments may be relied upon to indicate true power withing reasonably small limits of error.