首页   按字顺浏览 期刊浏览 卷期浏览 New Method for the Estimation of Lean Body Mass in Infants (EMME Instrument)V...
New Method for the Estimation of Lean Body Mass in Infants (EMME Instrument)Validation in Nonhuman Models

 

作者: W.,   Klish G.,   Forbes A.,   Gordon W.,  

 

期刊: Journal of Pediatric Gastroenterology and Nutrition  (OVID Available online 1984)
卷期: Volume 3, issue 2  

页码: 199-204

 

ISSN:0277-2116

 

年代: 1984

 

出版商: OVID

 

关键词: Lean body mass;EMME;Body conductivity

 

数据来源: OVID

 

摘要:

An innovative, safe, and rapid method suitable for the determination of lean body mass in infants is described. The machine (EMME) operates on the principle that a biologic specimen in an electromagnetic field perturbs the field proportional to its conductivity which is dependent on the amount of electrolyte present. The physical characteristics of the EMME signal are described. Experiments done with phantom infants composed of electrolyte solution and corn oil showed a very linear relationship between fat-free volume and the natural log of the EMME signal (r = 0.97). A similar relationship was noted for ground beef (r = 0.91) and live rabbits (r = 0.99). The EMME instrument appears to hold promise as a rapid noninvasive means of determining lean body mass in infants.

 