A Pulsed NMR System for Nuclear Thermometry Below 2 K

 

作者: M. I. Aalto,   H. K. Collan,   R. G. Gylling,   K. O. Nores,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1973)
卷期: Volume 44, issue 8  

页码: 1075-1082

 

ISSN:0034-6748

 

年代: 1973

 

DOI:10.1063/1.1686305

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A pulsed NMR system for platinum nuclear spin thermometry and for measurements of the nuclear spin‐lattice relaxation time &tgr;1has been developed. The nuclear free precession signal is detected with a fast linear detector and anRCintegrator. The &tgr;1measurements were performed by the transient recovery method with the aid of automated circuitry multiplexing of the linear detector output into four separate integrators. The thermometer was tested between 10 mK and 2 K using a powdered platinum sample. Within the experimental accuracy of about 2% the amplitude of the nuclear free precession signal was found to be proportional to 1/T. The thermometer could be calibrated with the aid of the Korringa relation &tgr;1T=&kgr;, with &kgr;=29.6 msec K. Thermal equilibrium problems of the &tgr;1measurements are discussed in detail.

 