首页   按字顺浏览 期刊浏览 卷期浏览 Spinal Opiate AdministrationA Case of Catheter Misplacement
Spinal Opiate AdministrationA Case of Catheter Misplacement

 

作者: Carol Tom,   Lorraine Arias,   Giancarlo Barolat,  

 

期刊: The Clinical Journal of Pain  (OVID Available online 1990)
卷期: Volume 6, issue 1  

页码: 60-63

 

ISSN:0749-8047

 

年代: 1990

 

出版商: OVID

 

关键词: Morphine pump;Catheter misplacement;Extra-arachnoid space

 

数据来源: OVID

 

摘要:

Abstract: Continuous spinal opiate administration via permanently implantable drug delivery devices has been proven to provide profound analgesia for chronic pain conditions. We present a case in which the catheter of an implantable subarachnoid device was misplaced into the subdural/extra-arachnoid space despite the free flow of cerebrospinal fluid. This was verified by x-ray dye studies. It is postulated that this misplacement of the catheter likely occurred as a result of recent lumbar punctures the patient had undergone. Extravasation of cerebrospinal fluid created a false space and contributed to the misplacement and ultimate failure of the device to provide analgesia

 

点击下载:  PDF (233KB)