Energy‐Independent Detector for Total Hydrogen Fluxes in the Range 1–10 keV for Space and Laboratory Applications
作者:
R. L. Wax,
W. Bernstein,
期刊:
Review of Scientific Instruments
(AIP Available online 1967)
卷期:
Volume 38,
issue 11
页码: 1612-1615
ISSN:0034-6748
年代: 1967
DOI:10.1063/1.1720617
出版商: AIP
数据来源: AIP
摘要:
An energy‐independent instrument capable of detecting total hydrogen fluxes (either H0or H0+H+) in the energy range 1–10 keV has been developed. Low extreme ultraviolet sensitivity has been achieved by scattering of the incident particles in a thin carbon foil to a detector located in an off‐axis position. Energetic electrons are removed by a magnetic field. Measurements of the ratio of the neutral to total fluxes leaving the foil, the angular distribution of the scattered particles, and the energy lost in transit of the foil are presented. The instrument has an efficiency for energetic hydrogen of 0.7%, an euv sensitivity of 2×10−3, and a geometrical factor of 0.06 cm2sr.