Calibration of microphones for sound intensity measurements

 

作者: A. F. Seybert,  

 

期刊: The Journal of the Acoustical Society of America  (AIP Available online 1986)
卷期: Volume 80, issue S1  

页码: 103-103

 

ISSN:0001-4966

 

年代: 1986

 

DOI:10.1121/1.2023511

 

出版商: Acoustical Society of America

 

数据来源: AIP

 

摘要:

The two‐microphone sound intensity method is potentially a valuable tool for noise source identification, sound power determination, and for other measurements (e.g., transmission loss) that are based on sound intensity measurements. An essential ingredient in an accurate sound intensity measurement is the accurate measurement of the phase between the two microphones. Because the phase is generally small, on the order of a few degrees, the microphones must either be “phase matched,” or a phase calibration procedure must be used. This paper outlines the design and test procedures for using a plane‐wave tube to perform a microphone phase calibration with an uncertainty of less than approximately 0.2 deg. The phase calibration procedure is illustrated for a pair of condenser microphones and for a pair of miniature crystal microphones typically used in hearing aids.

 