首页   按字顺浏览 期刊浏览 卷期浏览 Evaluation of the EMIT Amitriptyline and Nortriptyline Assays for the Determination of ...
Evaluation of the EMIT Amitriptyline and Nortriptyline Assays for the Determination of Serum Clomipramine and Desmethylclomipramine

 

作者: Albert Fraser,   Wendy Bryan,   Arthur Isner,  

 

期刊: Therapeutic Drug Monitoring  (OVID Available online 1989)
卷期: Volume 11, issue 3  

页码: 349-353

 

ISSN:0163-4356

 

年代: 1989

 

出版商: OVID

 

关键词: Antidepressants;Clomipramine;EMIT;High-performance liquid chromatography.

 

数据来源: OVID

 

摘要:

Summary: Homogeneous enzyme immunoassay reagents (EMIT) developed for the measurement of amitriptyline and nortriptyline in serum were modified to allow quantitation of clomipramine and desmethylclomipramine. The method was compared to a high-performance liquid chromatographic method. Between-run precision [coefficient of variation (CV)] for clomipramine in the EMIT assay for amitriptyline ranged from 2.6 to 3.2%. For desmethylclomipramine in the nortriptyline assay, the between-run CV ranged from 1.4 to 1.9%. Serum specimens from 43 patients (desmethylclomipramine) and 59 patients (clomipramine) were analyzed by both methods, with good correlation between methods. For clomipramine, recovery ranged from 100 to 102% (0–600 ng/ml range) and was 95–103% for desmethylclomipramine (0–600 ng/ml). The modified EMIT assays offered sufficient reproducibility, accuracy, and correlation with an established method for routine analysis of clomipramine and desmethylclomipramine.

 

点击下载:  PDF (317KB)



返 回