首页   按字顺浏览 期刊浏览 卷期浏览 Portable Blood Pressure MeasurementsPerformance of Korotkov Sound Analysis Techniques
Portable Blood Pressure MeasurementsPerformance of Korotkov Sound Analysis Techniques

 

作者: ROGER WOLTHUIS,   DAVID HULL,   DONALD MCAFOOSE,   JOSEPH FISCHER,  

 

期刊: Hypertension  (OVID Available online 1981)
卷期: Volume 3, issue 5  

页码: 596-600

 

ISSN:0194-911X

 

年代: 1981

 

出版商: OVID

 

关键词: blood pressure measurement;Korotkov sound;automatic BP measurement;portable BP measurement

 

数据来源: OVID

 

摘要:

SUMMARY Nonlnrasive portable blood pressure systems (PBPS) are typically based on the Korotkov sound technique. The performance of related Korotkov sound analysis techniques was assessed in PBPS data obtained from nine normotensive and nine hypertensive men. Multiple-seated PBPS measurements were taken, each following 1 minute of quiet rest, moderate walking, and stair climbing; each PBPS measurement was accompanied by a simultaneous auscultatory measurement in the same arm. In subsequent manual PBPS data analysis, PBPS systolic and diastolic BPs were within 4 mm Hg of corresponding auscultatory BPs 86% and 88% of the time; BPs following rest and walking showed greater accuracy, while those following stair climbing were less accurate. Interestingly, automated analysis of diastolic PBPS data (using selective bandpass filtering, amplitude normalization, and comparator decision ratios) provided a level of accuracy similar to that obtained from manual data analysis. Overall, both manual and automated Korotkov sound analysis techniques can provide BP results that agree quite well with auscultatory determinations.

 