A comparison of circular stapling devices for rectal anastomosesA study in dogs
作者:
Frank Keane,
Dermot Moore,
Patrick Byrne,
Thomas Hennessy,
期刊:
Diseases of the Colon & Rectum
(OVID Available online 1982)
卷期:
Volume 25,
issue 4
页码: 285-288
ISSN:0012-3706
年代: 1982
出版商: OVID
关键词: Rectum;Surgical staplers
数据来源: OVID
摘要:
&NA;Mechanical circular stapling devices offer several advantages in the formation of low rectal anastomoses. The author's aim was to compare the quality of rectal anastomoses created in dogs using the two commonly available staple suture guns, the EEA and SPTU. Thirty dogs, 15 with each gun, had rectal division and anastomosis. The first 20 dogs (ten EEA and ten SPTU) were sacrificed at eight days. Anastomotic integrity was assessed using plain x‐ray, barium enema, and postmortem examination and pressure studies. The remaining ten dogs, five in each group, were sacrificed at 48 hours, and the anastomoses were inspected and examined histologically. Eleven of 15 anastomoses in the EEA group showed evidence of breakdown compared with only three of 15 in the SPTU group (P<0.02). It is concluded that the SPTU gun created a more satisfactory anastomosis and that this was probably due to its staples having an adjustable closure.