首页   按字顺浏览 期刊浏览 卷期浏览 Rigid Lens Base Curve Stability upon Hydrogen Peroxide Disinfection
Rigid Lens Base Curve Stability upon Hydrogen Peroxide Disinfection

 

作者: MARCUS PICCOLO,   NORMAN LEACH,   ROGER BOLTZ,  

 

期刊: Optometry and Vision Science  (OVID Available online 1990)
卷期: Volume 67, issue 1  

页码: 19-21

 

ISSN:1040-5488

 

年代: 1990

 

出版商: OVID

 

关键词: rigid contact lenses;hydrogen peroxide;base curve;base curve warpage;contact lens disinfection

 

数据来源: OVID

 

摘要:

Because of the possibility of transmitting communicable diseases, in particular the HIV virus, it has been recommended that all diagnostic contact lenses, including rigid lenses, be disinfected after each use. Hydrogen peroxide is a recommended disinfection agent, but its effect on rigid lens polymers is relatively unknown. We soaked 50 lenses of 5 different polymers in 3% hydrogen peroxide for 10 min and measured the base curves to determine if any changes occurred. Our results showed no statistically significant change in base curve for the lenses measured, but 22% of the lenses did exhibit a small amount of warpage (mean 0.0382 mm) of the base curve. We were not able to determine what caused these lenses, and not all the lenses, to warp. Under the conditions of this study, hydrogen peroxide does not appear to cause clinically significant parameter changes, but lenses should be checked for warpage before use.

 