首页   按字顺浏览 期刊浏览 卷期浏览 Modified extractor gauge
Modified extractor gauge

 

作者: Y. Fujii,   H. Inoue,   H. Mima,   F. Kanematsu,  

 

期刊: Journal of Vacuum Science&Technology A: Vacuum, Surfaces, and Films  (AIP Available online 1983)
卷期: Volume 1, issue 1  

页码: 90-94

 

ISSN:0734-2101

 

年代: 1983

 

DOI:10.1116/1.572318

 

出版商: American Vacuum Society

 

关键词: design;performance;ultrahigh vacuum;vacuum gages;x radiation;pressure measurement

 

数据来源: AIP

 

摘要:

The design and performance of a modified extractor gauge for ultrahigh vacuum pressure measurement is described. The open end construction of the grid cage prevents the x rays from striking the ion collector directly except for those minor indirect x rays that are reflected at respective electrodes. The sensitivity factor is about 8.3×10−2Pa−1(11 Torr−1) at an electron current of 1 mA corresponding to the residual gases. Pressure measurements were carried out in the range of 1×10−6to 9×10−10Pa (1×10−8to 7×10−12Torr). It is confirmed that the x‐ray limit of the gauge is lower than that of the nude‐type B–A gauge. Description is also given of the modified design of the gauge for high vacuum pressure measurements facilitated by the cap electrode mounted on the top of the grid cage.

 

点击下载:  PDF (347KB)