首页   按字顺浏览 期刊浏览 卷期浏览 Micro-quantitation of lens myo-inositol by anion exchange chromatography
Micro-quantitation of lens myo-inositol by anion exchange chromatography

 

作者: DickersonJaime E.,   LouMarjorie F.,  

 

期刊: Current Eye Research  (Taylor Available online 1990)
卷期: Volume 9, issue 2  

页码: 201-205

 

ISSN:0271-3683

 

年代: 1990

 

DOI:10.3109/02713689008995207

 

出版商: Taylor&Francis

 

数据来源: Taylor

 

摘要:

A method capable of the determination of pmole quantities of myo-inositol in mgr amounts of tissue by anion exchange chromatography is detailed for use in lens and potentially other tissues. Samples were rendered protein-free through ZnS04/Ba(0H)2precipitation, lyophilized and reconstituted in water just prior to analysis. An aliquot of sample was injected onto an anion exchange column and eluted with a 0.045 M sodium hydroxide mobile phase. Each analysis requires 30 minutes to complete. Average recovery of myo-inositol added to lens sample prior to injection was 100.6%. The coefficient of variation for repeated sample injections was 2.9%. Rabbit lens averaged 11.4μmol/gr wet weight with epithelium containing 8.5μmol/gr wet weight while human lens contained 20.1μmol/gr wet weight and human epithelial cells had 17.5 nmol/mgr protein.

 

点击下载:  PDF (353KB)



返 回