An Evaluation of the Kodak Ektachem Clinical Chemistry Slide for Theophylline
作者:
P,
St. Louis M.,
MacDonald E.,
Giesbrecht S.,
期刊:
Therapeutic Drug Monitoring
(OVID Available online 1989)
卷期:
Volume 11,
issue 1
页码: 93-96
ISSN:0163-4356
年代: 1989
出版商: OVID
关键词: Theophylline;Kodak Ektachem slide;TDx.
数据来源: OVID
摘要:
We evaluated the Kodak Ektachem clinical chemistry slide for assay of theophylline. Assay precision and accuracy were acceptable in the therapeutic range although precision was poor at low levels of theophylline. The assay performed well with patients' samples using the Abbott TDx as the reference procedure but, as indicated by the manufacturer, uremic samples gave a positive bias, particularly in the therapeutic range. Finally, the significant bias observed with Quality Control material, probably due to matrix sensitivity, is a possible drawback.