首页   按字顺浏览 期刊浏览 卷期浏览 Calibration of Circular Dichrometers
Calibration of Circular Dichrometers

 

作者: KarlH. Pearson,   VictorC. Zadnik,   JamesL. Scott,  

 

期刊: Analytical Letters  (Taylor Available online 1979)
卷期: Volume 12, issue 9  

页码: 1049-1061

 

ISSN:0003-2719

 

年代: 1979

 

DOI:10.1080/00032717908059784

 

出版商: Taylor & Francis Group

 

关键词: Circular dichrometer calibration;Circular dichroism;Optical rotatory dispersion;Molar absorptivity

 

数据来源: Taylor

 

摘要:

The tris (hydroxymethyl)aminomethane salt of d-10-camphorsulfonic acid (TRIS-CSA) provides a convenient, non-hygroscopic alternative to the parent compound in the calibration of circular dichrometers. The molecular ellipticities, [6], of TRIS-CSA at 291 and 191 nm are 8015± 11 and -16, 130± 40 deg cm2dmol−1, respectively. The circular dichroism (CD) maxima in terms of δϵ are 2.429 1 cm−1mol−1at 291 nm and -4.888 1 cm−1mol−1at 191 nm. TRIS-CSA has an ORD peak at 306.5 nm, a zero crossover at 292 nm, and a trough at 270.5 nm (molecular rotation, [M]306.5=4475± 10, [M]270.5= -5615± 10 deg cm2dmol−1), and a molar absorptivity, ϵ, of 35.9± 0.2 1 cm−1mol−1at 286 nm.

 

点击下载:  PDF (308KB)



返 回