首页   按字顺浏览 期刊浏览 卷期浏览 Bestimmung der Empfindlichkeit bei Tc-99-NMR-Messungen an einem 250-MHz-Spektrometer
Bestimmung der Empfindlichkeit bei Tc-99-NMR-Messungen an einem 250-MHz-Spektrometer

 

作者: M. Findeisen,   B. Lorenz,   M. Wahren,  

 

期刊: Isotopenpraxis Isotopes in Environmental and Health Studies  (Taylor Available online 1990)
卷期: Volume 26, issue 11  

页码: 520-523

 

ISSN:0021-1915

 

年代: 1990

 

DOI:10.1080/10256019008622419

 

出版商: Taylor & Francis Group

 

关键词: accuracy;computers;nuclear magnetic resonance;quantity ratio;technetium compounds;technetium 99;sensitivity;signal-to-noise ratio

 

数据来源: Taylor

 

摘要:

By means of different concentrations the signal-to-noise ratios of Tc-99-NMR spectra were determined applying well measurable Tc-samples [tetrabutylammoniwn pertechnetate, TBA (TcO4)J and a 250 MHz-spectrometer. The signal-to-noise ratios of the spectrawere determined by using the integrated routines of the firm's software and accumulating different number of scans. By fittings of data of the signal-to-noise ratio dependence and by extrapolation the minimum Tc-concentralion could be empirically found out. Applying a duration of measurements of 12 hours about 10−7molar concentrations can be determined.

 

点击下载:  PDF (326KB)



返 回