首页   按字顺浏览 期刊浏览 卷期浏览 Postal Dose TLD Intercomparisons for Cobalt 60 Teletherapy Units in Bulgaria (1975 ...
Postal Dose TLD Intercomparisons for Cobalt 60 Teletherapy Units in Bulgaria (1975—1980)

 

作者: V. Penchev,   B. Constantinov,   K. Ivanova,  

 

期刊: Isotopenpraxis Isotopes in Environmental and Health Studies  (Taylor Available online 1983)
卷期: Volume 19, issue 3  

页码: 81-85

 

ISSN:0021-1915

 

年代: 1983

 

DOI:10.1080/10256018308544844

 

出版商: Taylor & Francis Group

 

关键词: accuracy;gamma dostinetry;IAEA;intercomparison;lithium fluorides;sensitivity;thermolumiuescent dosimetry;therapy;WHO

 

数据来源: Taylor

 

摘要:

During the period 1975—1980 in Bulgaria 6 TLD postal dose intercomparisons including all teletherapy units have been performed. LiF powder TLD-100 and a Harshaw TL analyser model 2000 have been used. The fading effect and the accuracy of the dose measurements for60Co gamma rays have been investigated preliminarily. The fading is below 1% for 30 days. The sensitivity of the reading equipment has been investigated using the mounted check light source. By means of a statistical analysis it was determined that the relative uncertainty due to unstabilities in the sensitivity is ±0,24% at confidence level P = 0,95 for a set of 10 successive check source readings. In the same manner the relative uncertainty of the dose measurements using one TL detector (4 measurements) was determined to be ±1,6% at P = 0,95. The intercomparison performance method is described and the results from the six intercomparisons are presented.

 

点击下载:  PDF (464KB)



返 回