Chapter 3: Practical Biomechanics for the Application of Ilizarov Fixators to Fractures of the Tibia
作者:
James Hutson,
期刊:
Techniques in Orthopaedics
(OVID Available online 2002)
卷期:
Volume 17,
issue 1
页码: 15-25
ISSN:0885-9698
年代: 2002
出版商: OVID
关键词: Ilizarov fixator;Biomechanics;Ring sizing
数据来源: OVID
摘要:
SummaryThe biomechanics of tensioned wires and half pins are described. The effects of wire position, number, and configuration on constructing a stiff periarticular fixation block are illustrated. Half pin application to the shaft and construction of a stable base for fixation of periarticular fractures are illustrated. The selection of appropriate ring diameter is based on soft tissue clearance and frame stiffness.