Nanosecond Time Resolution in Fission Fragment X‐Ray Lifetime Experiments Using Thin Plastic and NaI (Tl) Scintillation Detectors

 

作者: L. Bridwell,   M. E. Wyman,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1966)
卷期: Volume 37, issue 9  

页码: 1145-1150

 

ISSN:0034-6748

 

年代: 1966

 

DOI:10.1063/1.1720444

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Apparatus for the half‐life measurements of x rays from fission fragments is described. The fission fragment detector was an ultrathin plastic scintillator and the x‐ray detector was a 1 mm NaI (Tl) crystal. Both detectors were mounted on fast (56 AVP) photomultipliers. The details of the design of the plastic scintillator are given and curves showing the double humped fission fragment kinetic energy spectrum are illustrated. A systematic study of the x‐ray energy detection characteristics of the NaI (Tl) on a 56 AVP is given. The pulse heights were determined to be linear with x‐ray energy and the spectra were shown to follow Poisson's statistics. The number of photoelectrons produced at the photocathode was determined to be 3.2 per kiloelectron volt. The time measuring device was a conventional time to amplitude converter using pulse overlap methods. The prompt response of this converter could be predicted from the results gained in the energy detector study. Both the width and the apparent half‐life were consistent with the assumption that the time to amplitude converter was triggered on a single photoelectron.

 