Argonne 256‐Channel Pulse‐Height Analyzer
作者:
Robert W. Schumann,
James P. McMahon,
期刊:
Review of Scientific Instruments
(AIP Available online 1956)
卷期:
Volume 27,
issue 9
页码: 675-685
ISSN:0034-6748
年代: 1956
DOI:10.1063/1.1715670
出版商: AIP
数据来源: AIP
摘要:
A 256‐channel pulse‐height analyzer using a magnetic core memory is now in use at the Argonne National Laboratory. This analyzer uses a form of the Wilkinson method of generating numbers in response to input pulses. These numbers are proportional to the amplitudes of the input pulses, and correspond to the channel numbers into which the individual counts are to be recorded. The digital data processing circuits comprise a simple computer, with an internal memory which can store 256 sixteen digit binary numbers. A cathode‐ray tube display of the data in the form of a plot of countsversuschannel number is available during and after operation. Permanent readout is in the form of a similar curve traced by a recording potentiometer. Printed decimal readout is also available. The normal average analyzer dead time is 80 &mgr;sec, and pulse rates in excess of 5 million per minute at the input circuit can be tolerated without appreciable data distortion.