首页   按字顺浏览 期刊浏览 卷期浏览 Argonne 256‐Channel Pulse‐Height Analyzer
Argonne 256‐Channel Pulse‐Height Analyzer

 

作者: Robert W. Schumann,   James P. McMahon,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1956)
卷期: Volume 27, issue 9  

页码: 675-685

 

ISSN:0034-6748

 

年代: 1956

 

DOI:10.1063/1.1715670

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A 256‐channel pulse‐height analyzer using a magnetic core memory is now in use at the Argonne National Laboratory. This analyzer uses a form of the Wilkinson method of generating numbers in response to input pulses. These numbers are proportional to the amplitudes of the input pulses, and correspond to the channel numbers into which the individual counts are to be recorded. The digital data processing circuits comprise a simple computer, with an internal memory which can store 256 sixteen digit binary numbers. A cathode‐ray tube display of the data in the form of a plot of countsversuschannel number is available during and after operation. Permanent readout is in the form of a similar curve traced by a recording potentiometer. Printed decimal readout is also available. The normal average analyzer dead time is 80 &mgr;sec, and pulse rates in excess of 5 million per minute at the input circuit can be tolerated without appreciable data distortion.

 

点击下载:  PDF (756KB)



返 回