Distribution of Waste Anesthetic Gases in the Operating Room Air
作者:
Robert Piziali,
Charles Whitcher,
Rudolph Sher,
Robert Moffat,
期刊:
Anesthesiology
(OVID Available online 1976)
卷期:
Volume 45,
issue 5
页码: 495-500
ISSN:0003-3022
年代: 1976
出版商: OVID
关键词: Anesthetics;gases;trace concentrations;Equipment;exhaust systems;Operating rooms;air conditioning.
数据来源: OVID
摘要:
Epidemiologic and animal studies identify a relationship between chronic exposure to anesthetic gases and health hazards. Efforts to reduce exposure of personnel require an understanding of the distribution of anesthetic waste gases in the operating room air. Concentrations ol nitrous oxide and halothane were measured at numerous stations throughout an operating room and a delivery room in the absence of personnel. Air conditioning flow rates and flow patterns were varied, as was the height of the anesthetic gas source. Air flow patterns were found to dominate the anesthetic gas distribution, while buoyaney effects were neligible. Venting waste gases at the floor does not significantly reduce exposure of personnel. Areas of high concentration were observed; their occurrences and locations varied strongly with air flow patterns. The exhaust grille is the best location for a single measurement of the average room concentration. Recirculating air-conditioning systems reduce energy costs; however, only the non-recirculating portion of the air exchanges reduces waste gas concentrations.