Bestimmung des relativen Wirkungsgrades eines Ge(Li)-Detektors durch Messung einer standardisierten166mIIo-Lösung

 

作者: W. Heinig,   M. Rösseler,  

 

期刊: Isotopenpraxis Isotopes in Environmental and Health Studies  (Taylor Available online 1985)
卷期: Volume 21, issue 4  

页码: 136-140

 

ISSN:0021-1915

 

年代: 1985

 

DOI:10.1080/10256018508623467

 

出版商: Taylor & Francis Group

 

关键词: barium 133;calibration;efficiency;gamma detection;gamma spectroscopy;holmium 166

 

数据来源: Taylor

 

摘要:

In order to determine the relative efficiency of a Ge(Li) detector in the region of 8O–1000 keV, measurements of a standardized166mHo solution were performed. The radioactivity concentration of the166mHo solution was determined by measurements of a133Ba reference standard. The total number of calibration measurements was 120. The influence of experimental conditions such as sample manufacture, absorption of the vessels, source positions and dead time, on the accuracy of the measurements, was considered. The observed spectra were analyzed by a simple programme. In comparison with other calibration nuclides, the advantages of the multigamma-ray nuclide166mHo are simple manufacture, large half-life and homogeneous distribution of gamma-rays with absolute intensities greater than 5% in the region of 8O–1000 keV.

 