首页   按字顺浏览 期刊浏览 卷期浏览 Quick sample exchange manipulator for dual molecular beam and scanning tunneling micros...
Quick sample exchange manipulator for dual molecular beam and scanning tunneling microscopy experiments

 

作者: Daniel J. D. Sullivan,   Andrew C. Kummel,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1992)
卷期: Volume 63, issue 10  

页码: 4285-4288

 

ISSN:0034-6748

 

年代: 1992

 

DOI:10.1063/1.1143726

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Here we describe the design, construction, and operation of an ultrahigh‐vacuum manipulator for combined molecular beam and scanning tunneling microscopy experiments. The manipulator allows (1) sample transfer, (2) liquid‐nitrogen cooling, (3) electron beam heating, (4) rotation about the vertical and horizontal axes, and (5) electrical isolation of the sample. The unique features of the design are (a) direct transfer of a 3/8‐in.‐diam sample to a load‐lock sample carrier without the use of a second transfer arm and (b) sample mounting and electron beam heating systems that permit the sample to be several hundred degrees hotter than the sample mounting plate, thus minimizing sample contamination.

 