On the spatial and temporal resolution of ion temperature measurements based on neutral beam scattering: Comment on ‘‘Design of an ion temperature diagnostic based on neutral beam scattering’’ [Rev. Sci. Instrum.57, 1792 (1986)]
作者:
A. J. H. Donne´,
E. P. Barbian,
期刊:
Review of Scientific Instruments
(AIP Available online 1987)
卷期:
Volume 58,
issue 7
页码: 1320-1321
ISSN:0034-6748
年代: 1987
DOI:10.1063/1.1139412
出版商: AIP
数据来源: AIP
摘要:
Small‐angle neutral beam scattering is a promising technique to measure the local ion temperature in a hot plasma. We will show that in general a 10% accuracy in the temperature measurement can be obtained with a temporal resolution in the order of 1 ms and a spatial resolution of 10 cm. These values are less favorable than previous claims by Van Heeschetal.