High Pressure3He Gas Scintillation Neutron Detector
作者:
R. Lee Aamodt,
Leon J. Brown,
Gordon M. Smith,
期刊:
Review of Scientific Instruments
(AIP Available online 1966)
卷期:
Volume 37,
issue 10
页码: 1338-1340
ISSN:0034-6748
年代: 1966
DOI:10.1063/1.1719974
出版商: AIP
数据来源: AIP
摘要:
Construction details are given for a He gas scintillation neutron detector which operates to 238 atm at room temperature. A bakeable pressure‐vacuum manifold, on which the detector is evacuated and filled, also serves to remove contaminants from the DPS waveshifter. The detector appears stable when permanently sealed off. Curves are given for light output vs pressure, and some effects of impurities on such curves are noted. Measurements were made of relative sensitivity to60Co gammas and Pu‐Be neutrons. The gamma flux required to induce a current equal to that from 1 neutron/cm2sec is 15 MeV/cm2sec when the scintillating gas is4He, and 23 MeV/cm2sec when it is3He.