首页   按字顺浏览 期刊浏览 卷期浏览 Scanning tunneling microscope calibration and reconstruction of real image: Drift and s...
Scanning tunneling microscope calibration and reconstruction of real image: Drift and slope elimination

 

作者: V. Yu. Yurov,   A. N. Klimov,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 5  

页码: 1551-1557

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1144890

 

出版商: AIP

 

数据来源: AIP

 

摘要:

Drift, slope of sample, and indeterminate sensitivity of piezoceramics are considered as the origin of the linear scanning tunneling microscope (STM) image distortions. A special algorithm of STM scanning and reconstruction of image of unknown atomic structures is suggested and tested. This algorithm allows us to measure three components of the drift velocity and two angles, characterizing the average slope of scanning surface. On the one hand, using this algorithm, one can perform the STM calibration by a known surface structure (for example, the highly oriented pyrolytic graphite surface) even in the presence of a drift. This enables us to determine all the three piezoceramics constants of STM piezoscanner and the deviation of the real scanner axesXandYfrom orthogonality. On the other hand, using such a calibrated STM and the described algorithm, it is possible to obtain the real STM image and make measurements for unknown surfaces with atomic resolution without the distortions mentioned above. As it has been proved, the error of interatomic distance measurements does not exceed 5% (∼0.1 A˚) for any direction along the surface, even in the presence of a drift up to 0.4 A˚/s and the sample slope angle up to 10°.

 

点击下载:  PDF (978KB)