首页   按字顺浏览 期刊浏览 卷期浏览 In vivo versus in vitro polymerization of acrylic bone cement: Effect on material prope...
In vivo versus in vitro polymerization of acrylic bone cement: Effect on material properties

 

作者: W. L. Bargar,   S. A. Brown,   H. A. Paul,   T. Voegli,   Y. Hseih,   N. Sharkey,  

 

期刊: Journal of Orthopaedic Research  (WILEY Available online 1986)
卷期: Volume 4, issue 1  

页码: 86-89

 

ISSN:0736-0266

 

年代: 1986

 

DOI:10.1002/jor.1100040110

 

出版商: Wiley Subscription Services, Inc., A Wiley Company

 

关键词: Bone cement;Total joint replacement;Mechanical testing;In vivo model

 

数据来源: WILEY

 

摘要:

AbstractFlexural strength and molecular weight of polymethylmethacrylate bone cement were determined for cement polymerized in vivo in a canine total hip replacement and cement from the same batch polymerized in vitro. Both in vivo and in vitro samples showed a slight increase in flexural strength and modulus at 2 weeks, followed by a nearly equivalent decline at 8 weeks. In vitro and in vivo samples exhibit a comparable stress to failure, but the in vivo samples show a significantly increased strain and decreased modulus. Cement from the proximal region of the prosthesis has significantly decreased flexural strength, lower modulus, and lower molecular weight than cement from the distal region and plug, suggesting that failures in the proximal region of a prosthesis may be due to inferior physical and mechanical properties of the bone cement in that region.

 

点击下载:  PDF (355KB)



返 回