The interrelationship between intraocular pressure and Honan Intraocular Pressure Reducer pressure
作者:
GreenKeith,
HarmanDavid,
CheeksLisa,
期刊:
Current Eye Research
(Taylor Available online 1986)
卷期:
Volume 5,
issue 8
页码: 621-624
ISSN:0271-3683
年代: 1986
DOI:10.3109/02713688609015127
出版商: Taylor&Francis
数据来源: Taylor
摘要:
The relationship between the pressure applied to the enucleated human eye using the Honan Intraocular Pressure Reducer (HIPR) and the peak intraocular pressure as a function of initial intraocular pressure has been examined. The peak intraocular pressure is linearly related to the applied HIPR pressure whether the latter is 30, 50, or 75 mm Hg. The slopes relating peak intraocular pressure to initial intraocular pressure at different HIPR settings are parallel. Use of the HIPR at settings greater than 30 mm Hg and an initial IOP of greater than 30 mm Hg could compromise ocular vascular perfusion.