Cooled low‐noise preamplifier for a bolometer
作者:
H. P. Gush,
M. Halpern,
S. Knotek,
期刊:
Review of Scientific Instruments
(AIP Available online 1992)
卷期:
Volume 63,
issue 1
页码: 90-92
ISSN:0034-6748
年代: 1992
DOI:10.1063/1.1143776
出版商: AIP
数据来源: AIP
摘要:
A preamplifier has been built for use within a cryogenic system operating at a temperature of 2 K. The input stage is a 2N6483 JFET operating near 100 K. The equivalent input noise is less than 8 nV/Hz1/2at a frequency of 10 Hz.