Optimum event rate for a CCD detector

 

作者: B. D. Hall,   D. Reinhard,   R. Monot,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1995)
卷期: Volume 66, issue 3  

页码: 2668-2671

 

ISSN:0034-6748

 

年代: 1995

 

DOI:10.1063/1.1145607

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A simple event‐counting technique is considered that can be implemented on detector systems with adjustable exposure times. A single measurement cycle has two steps: exposure and readout. During readout a threshold is used to discriminate against background noise, making it impossible to differentiate between single or multiple events. The actual event rate can be estimated by accumulating the results of repeated measurement cycles, and applying a correction based on the probability for multiple events to occur. By considering the uncertainty in the estimation of the event rate, and assuming a Poisson process, it is shown that optimum exposure requires an average event count per exposure of roughly 1.7. The technique is applied to a high‐energy electron‐counting system based on a linear CCD photodetector. ©1995 American Institute of Physics.

 