首页   按字顺浏览 期刊浏览 卷期浏览 Heat capacity measurement of dielectric solids using a linear surface heater: Applicati...
Heat capacity measurement of dielectric solids using a linear surface heater: Application to ferroelectrics

 

作者: Seung‐Min Lee,   Sook‐ll Kwun,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1994)
卷期: Volume 65, issue 4  

页码: 966-970

 

ISSN:0034-6748

 

年代: 1994

 

DOI:10.1063/1.1144927

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A new ac calorimetry for dielectric solids was developed. A long and thin metal strip directly evaporated onto the surface of the sample is used as a heater and thermometer. Temperature oscillation of the heater is measured from the third harmonic voltage at the heater. Both the heat capacity and thermal conductivity of the sample are obtained by the frequency dependent phase and amplitude of the third harmonic voltage. Since heat capacity addenda and heat leaks in the measurement are extremely small, prior calibration is not required. The calorimeter is fully automated and has been tested on two ferroelectrics, KH2PO4and triglycine sulfate, at temperatures above 77 K. The optimal frequency region for the calorimetry is also discussed.

 

点击下载:  PDF (584KB)



返 回