Impurity level lifetime measurements using a lock‐in amplifier
作者:
Kenneth Edmondson,
Steve Agoston,
Radha Ranganathan,
期刊:
American Journal of Physics
(AIP Available online 1996)
卷期:
Volume 64,
issue 6
页码: 787-791
ISSN:0002-9505
年代: 1996
DOI:10.1119/1.18176
出版商: American Association of Physics Teachers
关键词: CHROMIUM;FLUORESCENCE;METASTABLE STATES;LIFETIME;LOCK−IN AMPLIFIERS;RUBY;ALUMINIUM OXIDES;RUBY LASERS;60.;01.50.P
数据来源: AIP
摘要:
Lifetimes of metastable levels of chromium in ruby are determined using an unusual technique. The fluorescence from the chromium is measured using an ac technique in which the excitation is modulated and the fluorescence is detected synchronously. The novelty lies in the measurement of the dependence of the signal on the modulation frequency and extracting the lifetime from the frequency dependence. The experiments and data analysis demand a detailed understanding of the methods of lock‐in amplification and synchronous detection and hence they can serve as a useful primer on the subject. The observed values for the lifetimes show excellent agreement with previous measurements.
点击下载:
PDF
(426KB)
返 回