Surface heparinization of central venous catheters reduces microbial colonization in vitro and in vivoResults from a prospective, randomized trial

 

作者: Pia MD Appelgren,   Ulrika MD Ransjo,   Lars MD Bindslev,   Frank MD Espersen,   Olle PhD Larm,  

 

期刊: Critical Care Medicine  (OVID Available online 1996)
卷期: Volume 24, issue 9  

页码: 1482-1489

 

ISSN:0090-3493

 

年代: 1996

 

出版商: OVID

 

数据来源: OVID

 

摘要:

ObjectiveTo evaluate in vitro and in vivo the efficacy of covalent end point-attached heparin to single-lumen polyurethane central venous catheters in reducing microbial adherence and colonization.DesignIn vitro study: A controlled bench study. In vivo study: A prospective, randomized, double-blind, clinical trial.SettingIntensive care unit in a 1200-bed teaching hospital.InterventionsIn vitro study: Adhesion of 17 radiolabeled clinical isolates of Staphylococci to catheters was examined in vitro. In vivo study: The outcome of heparinized and control catheters was compared in vivo in patients receiving long-term parenteral nutrition. Fifty-five adult patients were prospectively, blindly randomized to heparinized or control central venous catheters. The catheters, removed on clinical grounds, were analyzed with semiquantitative and quantitative cultures. Blood cultures were done at catheter removal.Measurements and Main ResultsIn vitro study: Coagulase-negative Staphylococci adhered less in vitro to heparinized catheters than to control catheters (p < .05). In vivo study: Among 32 central venous catheters, or patients who completed the study, catheter-associated bacteremia or fungemia was observed in five patients in the control group (n = 19) and in no patient with a heparinized catheter (n = 13) (p = .047). Four of 13 catheters in the heparin group were colonized compared with 14 of 19 in the control group (p = .03). Coagulase-negative Staphylococci were the most frequent microorganisms in both groups. The numbers of organisms found on colonized catheters were larger in the control group than in the heparin group.ConclusionsCovalent end point surface heparinization appears to have a great impact on both in vitro and in vivo bacterial colonization of central venous catheters. Such heparinization can be a practical and economical approach to the prevention of catheter-associated bacteremia or fungemia.(Crit Care Med 1996; 24:1482-1489)

 



返 回