DEVELOPMENT OF A NON-CORRODING RADIO-CHROMATOGRAPHIC SYSTEM AND MEASUREMENT OF ITS PARAMETERS
作者:
Ali Mardan,
J. Alstad,
J. O. Liljenzin,
期刊:
Solvent Extraction and Ion Exchange
(Taylor Available online 1990)
卷期:
Volume 8,
issue 1
页码: 117-135
ISSN:0736-6299
年代: 1990
DOI:10.1080/07366299008917989
出版商: Taylor & Francis Group
数据来源: Taylor
摘要:
A chromatograph with on-line gamma-detector compatible with corrosive eluants is described. The interplay among chromatographic resolution, speed of analysis, and precision of a radio-activity detector is reviewed. The precision of the peak area measurement of a radio-chromatographic peak can be improved by incorporating a multiscaler in the counting circuit. The void volume and cation exchange capacity of a column packed with experimental monodisperse 9.8 μm, highly (>40%) crosslinked porous resin particles have been measured by different methods. Measured cation exchange capacity by isotopic exchange between (23Na)r+ (22Na)sand the titration method are identical. The variation of overall height equivalent to a theoretical plate for the isotopic exchange with flow rate is reported. The extrapolated value of plate height at zero flow rate is about 3.2 times particle diameter.