Characteristics of Several Carbon Thermometer‐Heaters at Low Temperatures and in Magnetic Fields to 100 Kilogauss
作者:
E. W. Hornung,
D. N. Lyon,
期刊:
Review of Scientific Instruments
(AIP Available online 1961)
卷期:
Volume 32,
issue 6
页码: 684-687
ISSN:0034-6748
年代: 1961
DOI:10.1063/1.1717469
出版商: AIP
数据来源: AIP
摘要:
Several carbon thermometers have been prepared with various sizes of carbon particles and new rugged construction methods. The resistance of the thermometers has been measured as a function of temperature from 1.217° to 20.355°K and as a function of magnetic field from 0 to 94 500 gauss. No simple functional relationship holds between resistance and temperature or between resistance and magnetic field. An approximately square dependence of the resistance on the magnetic field holds for some thermometers at small fields, but in no case does this dependence continue at high fields.
点击下载:
PDF
(263KB)
返 回