Concordance and precision of dual X-ray absorptiometry with a 10 s scan

 

作者:

 

期刊: The British Journal of Radiology  (WILEY Available online 1996)
卷期: Volume 69, issue 825  

页码: 816-820

 

年代: 1996

 

DOI:10.1259/0007-1285-69-825-816

 

出版商: The British Institute of Radiology

 

数据来源: WILEY

 

摘要:

AbstractDevelopment of dual energy X-ray absorptiometry (DXA) scanners with multidetector array technology has resulted in greatly shortened scanning times. The Hologic QDR-4500 includes an ultrafast (10 s) “turbo” scan mode recommended by the manufacturer for fast screening studies or as an aid to positioning the patient prior to scanning using the normal fast (30 s), medium (1 min) or high definition (2 min) modes. The suitability of the turbo mode for use in routine clinical studies was assessed by examining the concordance of bone mineral density (BMD) measurements obtained in this mode with measurements obtained using the three normal scanning modes. Studies in 151 female patients showed statistically significant discrepancies in four out of the six scan sites studied with systematic differences of 2.9% and 3.1% being observed for the posteroantenor (PA) spine and intertrochanteric region of the hip, respectively.In vivoprecision for the 10 s scan found by performing duplicate measurements on 37 patients had a coefficient of variation of 1.3% for PA spine and 2.5% for femoral neck BMD. An investigation of the dependence of precision on body mass index (BMI) shows that the precision of spine and hip BMD was adversely affected with increasing BMI but the trend was statistically significant only in the spine. It was concluded that turbo mode scans are acceptable for routine clinical studies of the spine and hip but should not be used for longitudinal studies or patients with BMI greater than 30 kg m−2.

 

点击下载:  PDF (465KB)



返 回