首页   按字顺浏览 期刊浏览 卷期浏览 Comparison of a RP-HPLC Method with the Therapeutic Drug Monitoring System TDxfor the D...
Comparison of a RP-HPLC Method with the Therapeutic Drug Monitoring System TDxfor the Determination of Theophylline in Blood Serum

 

作者: I.N. Papadoyannis,   V.F. Samanidou,   H. Tsoukali,   F. Epivatianou,  

 

期刊: Analytical Letters  (Taylor Available online 1993)
卷期: Volume 26, issue 10  

页码: 2127-2142

 

ISSN:0003-2719

 

年代: 1993

 

DOI:10.1080/00032719308017457

 

出版商: Taylor & Francis Group

 

关键词: Theophylline;therapeutic drug monitoring system TDx;High-performance liquid chromatography;Comparison

 

数据来源: Taylor

 

摘要:

Blood serum samples of patients undergoing theophylline therapy were analysed in parallel with a new RP-HPLC method and the therapeutic drug monitoring TDx system. Both techniques are accurate, precise, technically easy, rapid, but the results taken by the TDx method are in all cases higher than HPLC. The mean increase is of about 56%.

 

点击下载:  PDF (408KB)