首页   按字顺浏览 期刊浏览 卷期浏览 The energy dispersive x‐ray detector: A quantitative model
The energy dispersive x‐ray detector: A quantitative model

 

作者: David C. Joy,  

 

期刊: Review of Scientific Instruments  (AIP Available online 1985)
卷期: Volume 56, issue 9  

页码: 1772-1779

 

ISSN:0034-6748

 

年代: 1985

 

DOI:10.1063/1.1138092

 

出版商: AIP

 

数据来源: AIP

 

摘要:

When a Si(Li) solid‐state energy dispersive detector is used to collect x‐ray photons with energies below 2 keV the characteristic peaks in the spectrum show significant deviations from the ideal Gaussian shape observed for higher energy photons. This effect is due to incomplete collection of the charge deposited by the incident photon in the diode. The magnitude of this effect, and the resultant distortion of the spectrum, can be computed by means of a Monte Carlo simulation which models the detector in terms of the three parameters which characterize it, junction depth, diffusion length, and surface recombination velocity. For values of these parameters typical of commercial detectors it is found that the incomplete charge correction can be as high as 30% for a line such as NK&agr;at 400 eV. By matching simulated and experimental spectral shapes, quantitative corrections for each x‐ray line of interest can be computed. The results suggest that alternative design strategies for detectors might permit a higher quality of performance to be achieved at low energies.

 

点击下载:  PDF (617KB)



返 回