首页   按字顺浏览 期刊浏览 卷期浏览 Oxygen transfer rates in a mammalian cell culture bioreactor equipped with a cell‐lift ...
Oxygen transfer rates in a mammalian cell culture bioreactor equipped with a cell‐lift impeller

 

作者: M. Johnson,   G. André,   C. Chavarie,   J. Archambault,  

 

期刊: Biotechnology and Bioengineering  (WILEY Available online 1990)
卷期: Volume 35, issue 1  

页码: 43-49

 

ISSN:0006-3592

 

年代: 1990

 

DOI:10.1002/bit.260350107

 

出版商: Wiley Subscription Services, Inc., A Wiley Company

 

数据来源: WILEY

 

摘要:

AbstractMeasurements ofkLa were carried out in 1. 5‐ and 5‐L New Brunswick Scientific CelliGen® bioreactors. The measuredkLa in water were identical for both vessel sizes operated in similar condition. The mass transfer rate increased with temperature, mixing speed, and aeration rate, with this last parameter being the most significant. Surface aeration alone gavekLa values of 0. 4 to 1. 6 h−1. A 25% decrease inkLa was observed above an aeration rate of 1. 6 vvm. This was caused by the particular foam breaker of the CelliGen bioreactor. Measurements ofkLa using a mammalian cell culture medium supplemented with 5% fetal calf serum (FCS) have confirmed the negative effect of the foam breaker onkLaThe measured value in this medium was 1. 2 h−1for all aeration rates, more than 60% of which was attributed to surface

 

点击下载:  PDF (483KB)



返 回