The Reproducibility of MTF Measurements
作者:
A.C. Marchant,
E.A. Ironside,
J.F. Attryde,
T.L. Williams,
期刊:
Optica Acta: International Journal of Optics
(Taylor Available online 1975)
卷期:
Volume 22,
issue 4
页码: 249-264
ISSN:0030-3909
年代: 1975
DOI:10.1080/713819048
出版商: Taylor & Francis Group
数据来源: Taylor
摘要:
MTF measurements have recently been carried out in a number of European laboratories on a typical wide-angle photographic lens. While showing smaller differences than those occurring in earlier intercomparisons, the results still exhibited a greater variation than is desirable if the MTF is to be adopted universally in the specification and assessment of lens performance. However, a detailed analysis of the results has indicated probable reasons for many of the differences, and when corrected in accordance with the results of this analysis the spread of MTF values has been greatly reduced, never exceeding ± 0·07 and usually amounting to less than ± 0·03. These residual differences are almost certainly smaller than anything which could be detected as a difference of picture quality.
点击下载:
PDF (690KB)