A cryostatic setup for the low‐temperature measurement of thermal diffusivity with the photothermal method
作者:
M. Bertolotti,
G. Liakhou,
R. Li Voti,
S. Paoloni,
C. Sibilia,
N. Sparvieri,
期刊:
Review of Scientific Instruments
(AIP Available online 1995)
卷期:
Volume 66,
issue 12
页码: 5598-5602
ISSN:0034-6748
年代: 1995
DOI:10.1063/1.1146026
出版商: AIP
数据来源: AIP
摘要:
A cryostatic setup is described to perform photothermal deflection measurements from room temperature to 77 K. The setup uses gaseous nitrogen as a medium where the photodeflection is produced. The ability of the system to work is demonstrated presenting some measurements of thermal diffusivity of high‐temperature superconductor samples and of yttrium‐iron garnets with variable aluminum content. ©1995 American Institute of Physics.