首页   按字顺浏览 期刊浏览 卷期浏览 Interference with negative particle sampling due to formation of positive ions in a qua...
Interference with negative particle sampling due to formation of positive ions in a quadrupole mass spectrometer

 

作者: D. G. Kuehn,   L. M. Chanin,  

 

期刊: Journal of Applied Physics  (AIP Available online 1980)
卷期: Volume 51, issue 12  

页码: 6411-6412

 

ISSN:0021-8979

 

年代: 1980

 

DOI:10.1063/1.327586

 

出版商: AIP

 

数据来源: AIP

 

摘要:

A quadrupole mass spectrometer has been utilized to study active discharges containing Cs vapor. Production of positive ions by electron impact within the mass spectrometer and subsequent secondary electron emission from cesiated surfaces produced spurious or pseudonegative ion signals which interfered with attempts to sample negative ions from the plasma. These effects may also occur in other gas discharge systems.

 