Method for using germanium thermometers in moderately high magnetic fields
作者:
A. Roy,
D. S. Buchanan,
D. M. Ginsberg,
期刊:
Review of Scientific Instruments
(AIP Available online 1985)
卷期:
Volume 56,
issue 3
页码: 483-485
ISSN:0034-6748
年代: 1985
DOI:10.1063/1.1138275
出版商: AIP
数据来源: AIP
摘要:
We have devised a simple method for extending the zero‐field calibration of a germanium resistance thermometer to include the effects of magnetic fields up to 5 T. We describe the application of this method to the use of a germanium thermometer at liquid‐helium temperatures. We outline a similar procedure to take into account the temperature variation of the calibration of a Hall probe.