Self Blood Glucose Monitoring: A Comparative Study of the Glucoscot/Glucopat And Reflolux/Haemo-Glukotest 20–800R System.
作者:
MoermanJ.,
Van CrombruggeP.,
SierensL.,
期刊:
Acta Clinica Belgica
(Taylor Available online 1985)
卷期:
Volume 40,
issue 5
页码: 285-291
ISSN:1784-3286
年代: 1985
DOI:10.1080/22953337.1985.11719095
出版商: Taylor&Francis
数据来源: Taylor
摘要:
SummaryTwo reflectance photometers, recently available in Belgium, were assessed for accuracy and precision in determining venous blood glucose.The Glucoscot/Glucopat system (A. Menarini, Diagnostics Department, Italy) was compared with the Reflolux/Haemo-Glukotest 20–800R system (Boehringer-Mannheim, West Germany). There was an excellent correlation between glucose concentrations by the laboratory method and Glucoscot (r = 0.9899) or Reflolux (r = 0.9928) reflectance photometer results.We found that we could visually read the Haemo-Glukotest 20–800R reagent strip accurately while with the Glucopat strip we underestimated glucose values higher than 250 mg/dl.Although clearly better results can be obtained by using a meter than by visually reading the strips we conclude that either of the assessed techniques can be used in blood glucose monitoring.