首页   按字顺浏览 期刊浏览 卷期浏览 Evaluation of Response Time of a Transcutaneous Oxygen Tension Electrode
Evaluation of Response Time of a Transcutaneous Oxygen Tension Electrode

 

作者: YOUSEF,   ABU-OSBA BRADLEY,   THACH ROBERT,  

 

期刊: Pediatric Research  (OVID Available online 1981)
卷期: Volume 15, issue 2  

页码: 143-146

 

ISSN:0031-3998

 

年代: 1981

 

出版商: OVID

 

关键词: arterial occlusion;PO2;response time

 

数据来源: OVID

 

摘要:

SummaryResponse characteristics of a transcutaneous PO2electrode to step changes in PaO2were determined.In vitrolag time and 95% response time of a transcutaneous PO2monitor were compared toin vivoresponse. Release of arterial occlusion was used to produce rapid local Pao2changes in human infants and adults.In vitrolag time and 95% response time varied according to whether an increment or decrement in PO2was produced.In vivo95% response time of this electrode was two to six times slower than previous estimates, and it varied significantly with the magnitude of the step change, subject age, anatomic location, and local hemodynamic factors. Also, we found thatin vivolag time is two to three times faster than previously reported.SpeculationUnder steady-state conditions, transcutaneous PO2correlates well with PaO2, but when rapid change is encountered as during oxygen administration or apnea, transcutaneous PO2will significantly underestimate PaO2change. On the other hand, the relatively short lag time suggests that the transcutaneous PO2electrode may be useful in following rapid PaO2changes qualitatively.

 

点击下载:  PDF (326KB)



返 回