首页   按字顺浏览 期刊浏览 卷期浏览 Characterization of Coulometric and Hypercoulometric ECD Mode of Operation
Characterization of Coulometric and Hypercoulometric ECD Mode of Operation

 

作者: P. Popp,   J. Leonhardt,  

 

期刊: Isotopenpraxis Isotopes in Environmental and Health Studies  (Taylor Available online 1983)
卷期: Volume 19, issue 5  

页码: 160-162

 

ISSN:0021-1915

 

年代: 1983

 

DOI:10.1080/10256018308544871

 

出版商: Taylor & Francis Group

 

关键词: electron capture;electron-capture detectors;environment;gas chromatography;space charge;radiation sources

 

数据来源: Taylor

 

摘要:

The coulometric mode of operation is given, if each sample molecule passing the ECD is converted into a negative ion and if each negative ion causes a decrease of the ionization current. The sensitivity of the detector in this case amounts to 96500 C/mol. A hypercoulometric response is given, if fragments from dissociative processes undergo electron capture at several times and if the enhancing effect of space charges is used. In this case the sensitivity of the ECD exceeds the upper limit given by Faraday's number. The operation conditions of both EGD-modes are described.

 