首页   按字顺浏览 期刊浏览 卷期浏览 A method to measure the MTF of digital x‐ray systems
A method to measure the MTF of digital x‐ray systems

 

作者: Richard A. Sones,   Gary T. Barnes,  

 

期刊: Medical Physics  (WILEY Available online 1998)
卷期: Volume 11, issue 2  

页码: 166-171

 

ISSN:0094-2405

 

年代: 1998

 

DOI:10.1118/1.595493

 

出版商: American Association of Physicists in Medicine

 

数据来源: WILEY

 

摘要:

A method has been devised to accurately measure the modulation transfer function (MTF) of digital x‐ray systems up to and, for undersampled systems, beyond the pixel Nyquist frequency ( fN). A phantom consisting of an array of parallel tungsten or similar wires is imaged, and discrete Fourier transforms of rows of pixel values are computed. Under suitable conditions of phantom orientation, wire diameter, wire spacing, and image magnification, the envelope of the modulus of the mean Fourier transform represents the system MTF. Experimental results extending beyond  fNare presented for an undersampled prototype digital chest x‐ray system and shown to be in reasonable agreement with predicted values. Employment of the method with other digital imaging modalities [i.e., computerized tomography (CT) scanners and nuclear magnetic resonance (NMR) units] is also discussed as well as error considerations and practical problems in implementing the method.

 

点击下载:  PDF (495KB)



返 回