For the first time Stark experiments in a pulsed nozzle Fourier transform microwave spectrometer are successfully performed with the pulsed molecular beam parallel to the resonator axis, together with a new geometry for the Stark electrodes which prevents line broadening due toE-field inhomogeneity. The coaxially oriented beam-resonator arrangement, introduced in the 1990s, is known to provide a considerable improvement in resolution and sensitivity in comparison to the perpendicular geometry, but Stark effect experiments performed in a microwave cavity, using the conventional cubic electrode geometry with this arrangement, suffer from an inhomogeneous field distribution. This gives rise to broadened profiles of the Stark shifted lines and a persistent zero-field line. The linewidth achieved with our new configuration is now close to the typical value found in the absence of a Stark field: less than 10 kHz at 10 GHz half width at half height, which makes the system very suitable for yielding precise values of the dipole moment, a piece of chemical information especially important in understanding both structure and charge distribution of molecular complexes. The performances of the novel design electrodes and the improvements achieved are demonstrated by several examples. ©1998 American Institute of Physics.