Deteemination of Total Chlorine Residual in Treated Wastewaters by Electrode
作者:
RichardL. Jenkins,
RodgerB. Baird,
期刊:
Analytical Letters
(Taylor Available online 1979)
卷期:
Volume 12,
issue 2
页码: 125-141
ISSN:0003-2719
年代: 1979
DOI:10.1080/00032717908055687
出版商: Taylor & Francis Group
关键词: potentiometry;chlorine electrode;wastewater analysis
数据来源: Taylor
摘要:
Total chlorine residuals of wastewater effluents as determined by an Orion 97–70 residual chlorine electrode and the phenylarsine oxide (PAO) back-titration were compared. At ≤1 ppm levels, the electrode and titrimetric methods compared favorably, but at>l ppm chlorine concentration the electrode was preferred. Use of the electrode for measuring chlorine residuals was limited to samples with a biochemical oxygen demand ≥50 mg/1 oxygen, while the PAO back-titration could be employed above this limit. The electrode was useful for field analysis; however, the electrode reading was strongly dependent upon temperature. An empirical graph was developed which allowed a post-analysis correction for the temperature differential between samples and standards.