首页   按字顺浏览 期刊浏览 卷期浏览 Application of the Iodide Ion-Selective Electrode in the Potentiometric Determination o...
Application of the Iodide Ion-Selective Electrode in the Potentiometric Determination of Chloral Hydrate

 

作者: M.T. M. Zaki,  

 

期刊: Analytical Letters  (Taylor Available online 1985)
卷期: Volume 18, issue 14  

页码: 1697-1702

 

ISSN:0003-2719

 

年代: 1985

 

DOI:10.1080/00032718508064471

 

出版商: Taylor & Francis Group

 

关键词: Iodide Ion-Selective Electrode;Potentiometric Determination;Chloral Hydrate

 

数据来源: Taylor

 

摘要:

A simple and sensitive micro method for chloral hydrate determination based on oxidation with iodine in chloroform solution is described. The produced iodide ion in the extract is determined using the iodide ion-selective electrode by either a direct measurement, standard addition technique or potentiometric titration with standard silver nitrate solution. Samples containing 0.1 - 4.0 mg chloral hydrate are analyzed with an average recovery of 99-9% and standard deviation of 0.1%.

 

点击下载:  PDF (179KB)



返 回