首页   按字顺浏览 期刊浏览 卷期浏览 Analysis of the reliability and validity of the kinetic communicator exercise device
Analysis of the reliability and validity of the kinetic communicator exercise device

 

作者: MARTIN FARRELL,   JAMES RICHARDS,  

 

期刊: Medicine and Science in Sports and Exercise  (OVID Available online 1986)
卷期: Volume 18, issue 1  

页码: 44-49

 

ISSN:0195-9131

 

年代: 1986

 

出版商: OVID

 

关键词: EXERCISE DEVICE;ISOKINETICS;KINCOM;REHABILITATION

 

数据来源: OVID

 

摘要:

The Kinetic Communicator (KINCOM), a microcomputer controlled, hydraulically driven device for the test, measurement, and rehabilitation of human joint function, was tested for the reliability and validity of its operating systems. Testing focused on the three primary functions of the KINCOM: lever arm position, lever arm velocity, and force measuring systems. Measurements of the functions were made via external devices and were compared to measurements simultaneously made by the KINCOM system. Tests were performed for both static and dynamic operation of the unit. Force measurements averaged a difference of 3.2% or less, lever arm speed was within 1.5% of the target speed in all cases, and no difference was found in position measurement. These discrepancies were felt to be due to calibration error and were not considered an indication of inaccuracy in the KINCOM system.

 